MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 40MM/+12; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 40MM/+12; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6260-9-440
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 10/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding pain involving a metal head was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information, including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.Lot specific voluntary recall v40 - recall - 2249697-08/29/2016-007r was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.The following devices were also listed in this report: trident 0 deg insert 40mm, cat#623-00-40g, lot#n3jmhe.Trident psl ha cluster 58mm, cat#542-11-58g, lot#17095402.Accolade plus tmzf hip stem #5, cat#6021-0537, lot#a62x2l.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
Dr.Removed a +12 the 40 metal head and x3 liner and replaced with a ceramic 7+5 head and eccentric liner.Implants that were not explanted: 542-11-58g, lot n3jmhe.6021-0537, lot 17095402.Update 17/october/2018: as reported in how was issue noticed: patient presented with pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40 COCR LFIT HEAD 40MM/+12
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8059788
Report Number0002249697-2018-03686
Device Sequence Number2024897
Product Code LPH
UDI-Device Identifier07613327032772
UDI-Public07613327032772
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K061434
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2012
Device Catalogue Number6260-9-440
Device Lot NumberRALMDD
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2018
Initial Report FDA Received Date11/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2249697-08/29/2016-007R
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age78 YR
Patient Weight95
-
-