| Catalog Number |
362800 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported with the use of the bd vacutainer® ppt¿ plasma preparation tube k2e 15.8 mg there was an issue with missing label.There was no report of injury or medical intervention.
|
| |
|
Event or Problem Description
|
|
It was reported with the use of the bd vacutainer® ppt¿ plasma preparation tube k2e 15.8 mg there was an issue with missing label.There was no report of injury or medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received samples and photos from the customer facility for investigation.The photos were evaluated and the customer¿s indicated failure mode for missing label with the incident lot was observed.Additionally, evaluation of the customer samples was performed and missing label was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Based on evaluation of the customer photos, the customer¿s indicated failure mode for missing label with the incident lot was observed.Additionally, evaluation of the customer samples was conducted and missing label was observed.Based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Search Alerts/Recalls
|