Catalog Number 362800
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported with the use of the bd vacutainer® ppt¿ plasma preparation tube k2e 15.8 mg there was an issue with missing label.There was no report of injury or medical intervention.
 
Event or Problem Description
It was reported with the use of the bd vacutainer® ppt¿ plasma preparation tube k2e 15.8 mg there was an issue with missing label.There was no report of injury or medical intervention.
 
Additional Manufacturer Narrative
Investigation summary: bd received samples and photos from the customer facility for investigation.The photos were evaluated and the customer¿s indicated failure mode for missing label with the incident lot was observed.Additionally, evaluation of the customer samples was performed and missing label was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Based on evaluation of the customer photos, the customer¿s indicated failure mode for missing label with the incident lot was observed.Additionally, evaluation of the customer samples was conducted and missing label was observed.Based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PPT¿ PLASMA PREPARATION TUBE K2E 15.8 MG
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key8059945
Report Number1917413-2018-03753
Device Sequence Number276982
Product Code JKA
UDI-Device Identifier50382903628006
UDI-Public50382903628006
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K972075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2019
Device Catalogue Number362800
Device Lot Number8061548
Initial Date Received by Manufacturer 10/23/2018
Supplement Date Received by Manufacturer10/23/2018
Initial Report FDA Received Date11/12/2018
Supplement Report FDA Received Date12/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-