Catalog Number 5550-G-319
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646)
Date of Event 07/21/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event was reported for the alleged loosening and pain involving an triathlon patella was reported.The event was not confirmed by medical review.Method & results: product evaluation and results: visual, dimensional, functional inspection, and material analysis were not performed as the item was not returned.Clinician review: a review of the provided medical records and x-rays by a clinical consultant indicated: ¿there is no documented follow-up subsequent to (b)(6) 2018 or evidence of complaints referable to the right knee except the occasional painless "knocking" sensation noted by the patient on (b)(6) 2018 but not noted in the physical examination.Knocking after total knee arthroplasty could relate to patellar tracking or contact with the ts post of the tibial component.It could also be soft tissue related or related to the retained staple in the femoral condyle.There is no evidence that factors associated with implant manufacturing or materials are responsible for this clinical situation or if it still exists.Should additional information become available, this report will be updated.¿ product history review: a review of the product history records could not be performed as the device was not properly identified.Complaint history review: a complaint history review could not be performed as the device was not properly identified.Conclusions: the reported event regarding the alleged loosening and pain was not confirmed nor the root cause determined.As stated in the clinician¿s review, ¿on (b)(6) 2018 an office visit note states, "six months status-post right tkr.Doing quite well.No pain, playing tennis.Feels a knocking in his knee occasionally.X-ray: excellent position, no problems.0° to 125°.Impression: doing well, return one year." for the noted ¿feels a knocking in his knee occasionally¿, the clinician noted, ¿there is no documented follow-up subsequent to (b)(6) 2018 or evidence of complaints referable to the right knee except the occasional painless "knocking" sensation noted by the patient on (b)(6) 2018 but not noted in the physical examination.Knocking after total knee arthroplasty could relate to patellar tracking or contact with the ts post of the tibial component.It could also be soft tissue related or related to the retained staple in the femoral condyle.There is no evidence that factors associated with implant manufacturing or materials are responsible for this clinical situation or if it still exists.Should additional information become available, this report will be updated.¿ device not available.
 
Event or Problem Description
Patient reported he is experiencing loosening and pain after his right knee surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON SYMMETRIC X3 PATELLA
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8060113
Report Number0002249697-2018-03693
Device Sequence Number1327092
Product Code MBH
UDI-Device Identifier07613327014761
UDI-Public07613327014761
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number5550-G-319
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2018
Initial Report FDA Received Date11/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age62 YR
Patient Weight78
-
-