US SURGICAL PUERTO RICO ENDO SHEARS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 176643
Medical Device Problem Code Melted (1385)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic cholecystectomy, the surgeon noticed melting on the shaft of the device after its use in the procedure.The case was finished when it was noticed.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A burnt area on the shaft of the device was burnt.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur when device is improperly handled where the shaft of the device contacts the uninsulated portion of an energized cautery device or conductive material.Damage to the shaft may compromise jaw actuation.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO SHEARS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key8060156
Report Number2647580-2018-05451
Device Sequence Number1697029
Product Code GEI
UDI-Device Identifier10884523000733
UDI-Public10884523000733
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K951589
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model Number176643
Device Catalogue Number176643
Device Lot NumberP8G0267X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2018
Initial Date Received by Manufacturer 10/17/2018
Supplement Date Received by Manufacturer03/13/2019
Initial Report FDA Received Date11/12/2018
Supplement Report FDA Received Date04/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age44 YR
-
-