PLEXUS MANUFACTURING SDN. BHD MYCARELINK; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number 24950G
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient saw a "goopy" substance under the remote monitor and it was alleged that it was the battery that was leaking.No troubleshooting taken.The monitor is expected to be replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product analysis: model: 24950g, serial/lot#: (b)(4).The remote monitor was returned and analysis revealed, a defective radio frequency (rf) head battery, radio frequency (rf) returned with nickel contacts.The remote monitor failed during interrogate test.Found error code 2300 in failure analysis (fa) log.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The monitor was replaced and returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYCARELINK
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key8060174
Report Number3004593495-2018-01081
Device Sequence Number11087810
Product Code NVZ
UDI-Device Identifier00643169552104
UDI-Public00643169552104
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number24950G
Device Catalogue Number24950G
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/07/2018
Supplement Date Received by Manufacturer04/08/2019
Initial Report FDA Received Date11/12/2018
Supplement Report FDA Received Date04/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
-
-