• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LEG HOLDER SURGICAL ASSISTANT(P/N 10200); ACCESSORIES, OPERATING-ROOM, TABLE (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LEG HOLDER SURGICAL ASSISTANT(P/N 10200); ACCESSORIES, OPERATING-ROOM, TABLE (KIT) Back to Search Results
Model Number 7209451
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/19/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device is not stay locked.The procedure was completed with no significant delay and no patient injuries.
 
Additional Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and it was observed the cross bar was not returned and the unit had been disassembled and the t-bar had been removed.Further evaluation observed the pinch plate was warped.The unit was reassembled and a functional test revealed that the device could not lock the t-bar position.The complaint was confirmed and the root cause was associated with mechanical component failure.
 
Event or Problem Description
It was reported that the device is not staying locked.The procedure was completed with no significant delay.No backup device was available.No patient injuries were reported.
 
Additional Manufacturer Narrative
The initial incident report was submitted with incorrect manufacturing site information and registration number (b)(4) ((b)(4)the correct manufacturing site information and registration number is (b)(4) ((b)(4)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEG HOLDER SURGICAL ASSISTANT(P/N 10200)
Common Device Name
ACCESSORIES, OPERATING-ROOM, TABLE (KIT)
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048 6512
MDR Report Key8060473
Report Number1643264-2018-00859
Device Sequence Number11152352
Product Code FWZ
UDI-Device Identifier03596010468635
UDI-Public03596010468635
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7209451
Device Catalogue Number7209451
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2018
Initial Date Received by Manufacturer 10/23/2018
Supplement Date Received by Manufacturer12/14/2018
12/14/2018
Initial Report FDA Received Date11/12/2018
Supplement Report FDA Received Date12/18/2018
02/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-