| Model Number |
7209451 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device is not stay locked.The procedure was completed with no significant delay and no patient injuries.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and it was observed the cross bar was not returned and the unit had been disassembled and the t-bar had been removed.Further evaluation observed the pinch plate was warped.The unit was reassembled and a functional test revealed that the device could not lock the t-bar position.The complaint was confirmed and the root cause was associated with mechanical component failure.
|
| |
|
Event or Problem Description
|
|
It was reported that the device is not staying locked.The procedure was completed with no significant delay.No backup device was available.No patient injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial incident report was submitted with incorrect manufacturing site information and registration number (b)(4) ((b)(4)the correct manufacturing site information and registration number is (b)(4) ((b)(4)).
|
| |
|