WRIGHTS LANE SYNTHES USA PRODUCTS LLC CANN CONNECTING SCR W/INTERNL THRD F/PERCUTAN INSERTN HANDLE; GUIDE, SURGICAL, INSTRUMENT
|
Back to Search Results |
|
| Catalog Number |
03.010.146 |
| Medical Device Problem Code |
Device Difficult to Maintain (3134)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, product was not returned.Device was used for treatment, not diagnosis.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, patient underwent an orthopedic procedure.During the procedure, the surgeon inserted a tibial expert nail instrumentation and at the end of the procedure, he had a very difficult time removing the cannulated connecting screw with internal thread for percutaneous insertion handle from the nail.The knee was flexed but the screw was extremely hard to turn.The surgeon then asked that both the insertion handle bolts should be replaced in the set.There was no patient consequence reported.It is unknown if there was a surgical delay.Patient and procedure outcome were unknown.This report is for one unknown (1) connecting scr w/internal third f/percutan insertn handle.Concomitant device reported: unknown insertion handle (part#: unknown, lot#: unknown, quantity#: 1).This complaint involves one (1) device.This report is 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part: 03.010.146; synthes lot: u180796; supplier lot: n/a; release to warehouse date: july 02, 2013; expiration date: n/a; supplier: orchid unique.No nonconformance reports (ncrs) were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Flow: device interaction/functional.Visual inspection: the cannulated connecting screw w/internal third f/percutaneous insertion handle was received at us cq with minimal signs of wear.Functional test: a functional test could not be completed since the connecting screw was returned by itself and not with the other products mentioned in this complaint.The complaint could not be replicated; therefore the complaint is not confirmed.Dimensional inspection: dimensional analysis was completed.The inner diameter of the threaded portion of the connecting screw, measured 4.60 mm.This is within specification of 4.60 mm ± 0.1.The outer diameter on the other end of the connecting screw measured 12.20 mm which is within specification of 12.0 mm to 12.3 mm.The threaded portion of the connecting screw dimension mj8x1.25 ¿ 4h6h (7.788 to 8.000) measured 7.85 mm.The hex across flats measured 8.10 mm which is within specification of 8.025 mm to 8.115 mm.All measurements were taken using caliper 802 and based on drawing.Document/specification review: the following drawing(s) was reviewed: cannulated connecting screw with internal m6x1 thread for 03.010.046.Conclusion: the complaint condition is not confirmed as the device (part # 03.010.146, lot # u180796) was received with minimal signs of wear.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined, it is possible that the device encountered unintended forces.No new malfunctions were observed during the course of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|