BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U35130620
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was not provided, a manufacturing review could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly had difficulty retracting through the sheath.No further treatment was needed.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned.A visual inspection found the device stuck in the user's introducer sheath.The balloon was protruding from the sheath tip, and the tip of the sheath was buckled and flared.Therefore, the investigation is confirmed for a sheath-related retraction issue.The definitive root cause for the balloon becoming stuck in the sheath could not be determined based upon available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly had difficulty retracting through the sheath.No further treatment was needed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8060852
Report Number2020394-2018-02003
Device Sequence Number13950050
Product Code LIT
UDI-Device Identifier00801741093234
UDI-Public(01)00801741093234
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2021
Device Model NumberU35130620
Device Catalogue NumberU35130620
Device Lot NumberCMCP0076
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2018
Initial Date Received by Manufacturer 10/16/2018
Supplement Date Received by Manufacturer01/25/2019
Initial Report FDA Received Date11/12/2018
Supplement Report FDA Received Date02/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-