| Model Number |
989605350452 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported an articulation issue with their x7-2t model transducer.There was no injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A thorough evaluation of the x7-2t model transducer identified damage to the device.Visual inspection of this probe revealed wear to the cable, i-tube, and window.Further performance testing found the steering cable ferrule had insufficient solder resulting in the failure to articulate in the anterior direction.An evaluation for potential design change improvements are currently underway.
|
| |
|