• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. TRICUT® STRAIGHT SHAFT 4MM 13CM LONG M4 ROTATABLE WITH EM TRACKING; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. TRICUT® STRAIGHT SHAFT 4MM 13CM LONG M4 ROTATABLE WITH EM TRACKING; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884080EM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis found that visually, the distal tip had broken off the inner cutter.The portion that became detached measured approximately 0.24¿ in length.The break occurred at the first proximal tooth valley.The assemblies were deformed in such a way that indicates the inner blade teeth impacted the outer cutting edge at the 3rd proximal valley while moving in a counterclockwise direction which resulted in the observed break.There was no evidence of concentricity issues or bends in the returned tube / shaft assemblies.There was uneven wear at the cutting tip as well.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported via a manufacturer representative that the tip of the blade broke off.There was no patient impact.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Upon follow-up, it was mentioned that the event happened during a sinus surgery.The case was completed with a replacement device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICUT® STRAIGHT SHAFT 4MM 13CM LONG M4 ROTATABLE WITH EM TRACKING
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key8061182
Report Number1045254-2018-00588
Device Sequence Number1902500
Product Code EQJ
UDI-Device Identifier00613994617422
UDI-Public00613994617422
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K130608
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/11/2020
Device Model Number1884080EM
Device Catalogue Number1884080EM
Device Lot Number0215721480
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2018
Supplement Date Received by Manufacturer12/18/2018
Initial Report FDA Received Date11/12/2018
Supplement Report FDA Received Date01/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-