| Catalog Number |
014396 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the wire broke off.Broken piece was removed from patient.No patient injury or delay was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One 2.4 drill tip guide wire was returned for evaluation.Visual assessment of the guide wire confirmed the reported breakage.The guide wire is bent at the break area.A major area of abrasion is located at the breakage.The condition of the device indicates the drill came in contact with the guide wire when over drilling causing the observed damage.Drilling over a bent guide wire would require the use of excessive force.Per the devices ifu ¿as with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in failure¿.
|
| |
|