• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. NIM-ECLIPSE® STIMULATOR EXTENDER; STIMULATOR, ELECTRICAL, EVOKED RESPONSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. NIM-ECLIPSE® STIMULATOR EXTENDER; STIMULATOR, ELECTRICAL, EVOKED RESPONSE Back to Search Results
Model Number 945EEX901
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 945eex901, serial/lot #: (b)(4)/215046297, udi#: (b)(4) ; product id: 945daq916, serial/lot #: (b)(4)/212935723, udi#: (b)(4).Analysis on serial number #(b)(4) found broken cosmetic nuts.The cosmetic nuts were replaced.The device was repaired, tested, and passed manufacturing specifications.Analysis on serial number #(b)(4) found a loose connector port and a missing cosmetic nut.The cosmetic nut was replaced.The device was repaired, tested, and passed manufacturing specifications.Analysis on the amplifier found a defective front-end board and middle board.The affected parts were replaced.The device was repaired, tested, and passed manufacturing specifications.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care professional (hcp) reported that the amplifier and stimulators were not working.There was no known patient impact reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-ECLIPSE® STIMULATOR EXTENDER
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key8061505
Report Number1045254-2018-00589
Device Sequence Number13997988
Product Code GWF
UDI-Device Identifier00643169071865
UDI-Public00643169071865
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K050798
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number945EEX901
Device Catalogue Number945EEX901
Device Lot Number213083863
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2018
Initial Report FDA Received Date11/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-