The device and the lens were returned separated.The device was evaluated.The plunger is oriented correctly.There appears to be inadequate viscoelastic observed in the device.The plunger has been retracted to mid-nozzle.The nozzle tip has a large aneurysm on the bottom and a small tear on the side.This damage would indicate the lens/plunger were not in a proper position for advancement.The nozzle was removed and cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.The lens was loose in the plastic bag.Viscoelastic is observed on the lens.One haptic is broken-not returned.It is unknown if the haptic damage occurred during the delivery or the lens removal.The lens has been cut in half typical of a removal.One portion has a chipped area on the edge with material removed.A qualified viscoelastic was indicated.The root cause for the reported event could not be determined.The device tip exhibited damage (large aneurysm, and a small split) that would indicate the lens/plunger were not in acceptable positions for advancement.The dfu instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.A broken haptic was also observed.Top coat dye stain testing was conducted with acceptable results.In addition, there did not appear to be adequate viscoelastic in the device.The dfu instructs: fill the device until viscoelastic can be observed flowing to the line on the nozzle tip.This will require approximately 0.2 ml of viscoelastic.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device.The manufacturer internal reference number is: (b)(4).
|