| Model Number |
6707-35 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/23/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was returned and analyzed but no issue was identified that required full analysis.Analysis was performed and no anomalies were found.Concomitant medical products: 0041 boston scientific lead, implanted: (b)(6) 1993; 383059 lead, implanted: (b)(6) 2010; 4047 boston scientific lead, implanted: (b)(6) 2012; 6721l-50 lead, implanted: (b)(6) 2012; d143 boston scientific icd, implanted: (b)(6) 2015.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the lead was explanted due to a malfunction of the connector.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis the distal portion of the lead was returned, analyzed.Analysis indicated that the right ventricular defibrillation coil developed a fracture due to flexing while in vivo.If information is provided in the future, a supplemental report will be issued.
|
| |
|