| Model Number |
DL-330-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/04/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The ecog and impedance data are suggestive of a potential lead break.Neuropace was unable to determine a root cause.The explanted product was not returned to neuropace, inc.
|
| |
|
Event or Problem Description
|
|
Based on the review of the patient's ecogs by the treating center, a signal artifact was identified on the right depth lead indicating a potential lead break.Per the treating center, the patient fell due to a seizure in may 2018 and hit her head which is felt to be the probable cause of the possible lead break.The patient does not frequently interrogate and synchronize her neurostimulator and therefore this issue was not identified until (b)(6) 2018.During a routine neurostimulator replacement on (b)(6) 2018, the lead with the signal artifact was explanted and a new lead was implanted and connected to the rns neurostimulator.
|
| |
|