| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
A third party supplier to a customer has reported that dbs electrodes, which have been implanted with the leksell vantage frame have shifted in a medial-posterior direction and a revision surgery is required.
|
| |
|
Event or Problem Description
|
|
A third party supplier to a customer has reported that dbs electrodes, which have been implanted with the leksell vantage frame have shifted in a medial-posterior direction and a revision surgery is required.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was performed at site by another supplier and it has been concluded that the vantage arc system was working as designed and intended.The customer will contact the supplier of the mri (not supplied by elekta) to investigate the cause of the problem.
|
| |
|