• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; POUCH, COLOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; POUCH, COLOSTOMY Back to Search Results
Model Number 022767
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event or Problem Description
It was reported that the precut starter holes in the mass were cut off center.No harm was reported and no photo was provided.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: d4: unique identifier (udi).H6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions.H10: investigation summary.Batch record review: lot 8d00477 was manufactured on 04/17/2018 in the guard line with a total of (b)(4) market units.Complaint investigator performed a batch record review on 02/08/2019 to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bom (bill of materials), ds10-034 (system 5 drainable ostomy pouch with stomahesive skin barrier), process instructions (pouch disc collar assembly system 5 b (guard machine), and all the tooling information documented was also correct, under international commodity code (icc) code 22767 system application product (sap) material identification (id) 1003062 and manufacturing order 1399408.The batch record review supports that there were no discrepancies related to the issue reported.Photo/sample analysis: no photos/samples were received related to the reported problem.Investigation conclusion: based on the investigation findings through revision of the batch records documentation, process observation, personnel interviewed, methodology implemented and defects simulation, the root cause could not be identified.Nevertheless, there is one (1) condition that could be considered, based on objective evidence, as a contributor factor for this event: there is not control in place for centering pins in the guard line.No issues related to machine, material, method, measurement or environment were identified during the manufacturing process evaluation, therefore no further actions are required related to these factors.The investigation associated with related event has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092.Manufacturing site: 9618003.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVELIFE
Common Device Name
POUCH, COLOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key8065316
Report Number9618003-2018-03339
Device Sequence Number1953679
Product Code EZQ
UDI-Device Identifier00768455179213
UDI-Public00768455179213
Combination Product (Y/N)N
Initial Reporter StateOH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/16/2023
Device Model Number022767
Device Lot Number8D00477
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2018
Supplement Date Received by Manufacturer11/20/2023
Initial Report FDA Received Date11/13/2018
Supplement Report FDA Received Date11/29/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexFemale
-
-