| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Dizziness (2194)
|
| Date of Event |
10/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) internal report # (b)(4) returned meter and returned test strips evaluated with no defects found.Most likely underlying root cause: mlc-62-user had poor technique test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for erratic blood glucose results.The customer is concerned with results obtained of 277 and 212mg/dl.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of feeling dizzy and in a fog.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the dining room.During the call on (b)(6) 2018, a back to back blood test was performed fasting by the customer and produced test results of 259 and 273mg/dl using true metrix meter.The test strip lot manufacturer's expiration date is 4/30/2020 and open vial date is three weeks ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly).(b)(6).
|
| |
|