| Model Number |
BIO-COMP INTER-FERENCE SCRW |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
10/24/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee surgery the screw was broken off.The broken part was not retrieved from the patient and remains in the bone.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update 31, oct, 2018: the broken fragment is tibially fixed in the bone and can not move.No x-ray is available.The broken part remains in the bone, no further measures were taken.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Visual evaluation revealed that the bio-composite screw had broken off at the fourth thread prior to receipt for investigation.Complainant's event is most likely caused by not inserting the implant co-axial to the bone tunnel, prying or leveraging, improper bone preparation or flexing the joint during insertion.Per product directions for use (dfu-0111), it is important to completely seat the screwdriver to prevent potential screw fracture during insertion.
|
| |
|