Model Number BIO-COMP INTER-FERENCE SCRW
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 10/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was requested/is expected but has not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that during a knee surgery the screw was broken off.The broken part was not retrieved from the patient and remains in the bone.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update 31, oct, 2018: the broken fragment is tibially fixed in the bone and can not move.No x-ray is available.The broken part remains in the bone, no further measures were taken.
 
Additional Manufacturer Narrative
Complaint confirmed.Visual evaluation revealed that the bio-composite screw had broken off at the fourth thread prior to receipt for investigation.Complainant's event is most likely caused by not inserting the implant co-axial to the bone tunnel, prying or leveraging, improper bone preparation or flexing the joint during insertion.Per product directions for use (dfu-0111), it is important to completely seat the screwdriver to prevent potential screw fracture during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMP INTER-FERENCE SCRW
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key8065748
Report Number1220246-2018-00743
Device Sequence Number19368366
Product Code MAI
UDI-Device Identifier00888867035331
UDI-Public00888867035331
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K071176
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBIO-COMP INTER-FERENCE SCRW
Device Catalogue NumberAR-5028C-09
Device Lot Number10191786
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2018
Initial Date Received by Manufacturer 10/24/2018
Supplement Date Received by Manufacturer10/24/2018
Initial Report FDA Received Date11/13/2018
Supplement Report FDA Received Date12/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-