| Catalog Number |
4004C1030 |
| Medical Device Problem Codes |
Material Deformation (2976); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2018-02234.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure using penumbra coil 400s (pc400s).During the procedure, the physician felt resistance while advancing a pc400 through a lantern delivery microcatheter (lantern), despite multiple attempts to advance it.The physician therefore removed the lantern and the pc400.Upon removal, the physician noticed that the pc400 was fractured and the lantern was kinked.The procedure was completed using a px slim delivery microcatheter (px slim) and a new pc400.There was no report of an adverse effect to the patient.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure using penumbra coil 400s (pc400s).During the procedure, the physician felt resistance while advancing a pc400 through a lantern delivery microcatheter (lantern), despite multiple attempts to advance it.The physician therefore removed the lantern and the pc400.It was reported that the tip of the lantern was kinked.Additionally, it was noted that distortion was confirmed at the tip and at the end of the embolization coil; however, no damage was recognized on the pusher assembly.The procedure was then completed using a px slim delivery microcatheter (px slim) and a new pc400.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Results: the pet lock was intact on the proximal end of the pusher assembly.The mid-joint of the pusher assembly was retracted past the introducer sheath friction lock.The embolization coil was intact with the distal detachment tip (ddt) of the pusher assembly.The embolization coil was compressed and had offset coil winds.During functional testing, the pc400 introducer sheath was flushed and was able to be retracted onto the pusher assembly and the embolization coil was able to advance out of the introducer sheath and through a demonstration lantern with minor resistance.Resistance was encountered at the ovalized location on the lantern and the returned pc400 was unable to advance any further; therefore, no further functional testing was performed.Conclusions: evaluation of the returned pc400 revealed a device with offset coil winds, but no fracture.The reported complaint mentioned that resistance was experienced at the distal end of the lantern microcatheter when advancing the pc400.If the pc400 is advanced against resistance, damage such as this may occur.The ovalization found on the distal tip of the lantern was determined to be the probable cause of the reported resistance experienced during the procedure.Evaluation of the returned lantern revealed a flattened distal tip.Possible causes of this damage could be forcefully gripping or pinching the device during use; or interaction with the stent at the location of the procedure.During functional testing, the returned pc400 was unable to advance past the distal ovalization on the lantern.The cause of the ovalization could not be determined.Penumbra coils and catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|