| Model Number |
42856 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported the nurse was attempting to adjust the valve with the handtools prior to opening the package, but the valve would not adjust past level 1.0.Multiple attempts were made to no avail.A different valve was used andworked fine.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that the valve was not adjusted on a metal table.Alignment was maintained between the adjustment tool and valve setting when the distance between the two was less than 15cm.When the pressure level was checked, the adjustment tool was at least 76cm away from the valve.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was not removed from the sealed sterile tyvek pouch.Laboratory personnel were able to adjust the performance level settings from 0.5 to 2.5 then down to 0.5 again though the packaging.Therefore, the nature of the complaint could not be replicated.Due to the nature of the complaint the valve was not tested for patency, leak, reflux, siphon, valve flux, pressure-flow and preimplantation.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|