The returned valve was not removed from the sealed sterile tyvek pouch.Laboratory personnel were able to adjust the performance level settings from 0.5 to 2.5 then down to 0.5 again though the packaging.Therefore, the nature of the complaint could not be replicated.Due to the nature of the complaint the valve was not tested for patency, leak, reflux, siphon, valve flux, pressure-flow and preimplantation.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|