| Catalog Number |
0385545000 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip broke during the procedure.No pieces fell in the joint and the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: inner sheath of blade, tip broke off during case the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be excessive pressure, such as bending or prying, may cause the arthroscopic shaver blades to bend/break.Blade comes in contact with a hard object causing damage to the teeth and hitting the housing edges.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip broke during the procedure.No pieces fell in the joint and the procedure was completed successfully.
|
| |
|