• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST; VESSEL SEALER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST; VESSEL SEALER Back to Search Results
Model Number 480322-07
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Based on the information provided, this event is being reported due to the following conclusion: during a da vinci-assisted surgical procedure, the vessel sealer instrument was not sealing properly.While there was no report of any patient, harm, adverse outcome or injury, recurrence of the reported failure mode could cause or contribute to an adverse event.
 
Event or Problem Description
It was reported that during a da vinci-assisted total colectomy procedure, the vessel sealer instrument was not sealing properly.The customer stated that an insufficient amount of energy was applied to the instrument tips.The customer stated that the "tissue bites" were not larger than normal and that the instrument gave indications that the sealing was complete but was not fully sealing the tissue.The vessel sealer instrument was used for approximately seven fires before it was swapped out to another vessel sealer instrument.However, the issue persisted and once again the instrument was swapped out.The procedure was completed with no report of patient harm, adverse outcome or injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the patient did not experience any bleeding, an error message was displayed before cutting, however, it is unknown what the error message was.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the vessel sealer involved with this complaint and completed the device evaluation.Failure analysis did not confirm the customer reported complaint.The instrument was placed and driven on an in-house system.The instrument passed the self-test, electrical continuity test, and recognized the erbe generator unit on multiple attempts.The instrument moved intuitively with full range of motion in all directions.The grips opened and closed properly, energy delivery test was performed and passed.Performed grip force test on grips as additional testing, and it measured at 6.28 lbf in straight orientation, 5.98 lbf in yaw, and 6.0 lbf for pitch.All readings were above the minimum requirement of 4.25 lbf.Electrode gap inspection was tested as an additional functional check.Electrode gap test passed.In addition, verified gap between grips was.002", which is within range of.001" and.007".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Common Device Name
VESSEL SEALER
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
950 kifer rd.
sunnyvale CA 94086
MDR Report Key8067197
Report Number2955842-2018-10669
Device Sequence Number19430486
Product Code NAY
UDI-Device Identifier10886874112745
UDI-Public(01)10886874112745(17)200630(10)M10180615
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K140189
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number480322-07
Device Lot NumberM10180615 0371
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Initial Date Received by Manufacturer 10/17/2018
Supplement Date Received by Manufacturer12/12/2018
Initial Report FDA Received Date11/13/2018
Supplement Report FDA Received Date12/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-