BARD ACCESS SYSTEMS SIRE-RITE 8 ULTRASOUND SYSTEM 32MM PROBE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770601Y
Device Problems No Display/Image (1183); Poor Quality Image (1408)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer, at this time, for evaluation.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.Device has not been received, at this time.
 
Event Description
Picc rn reported system's ultrasound image is very dark.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of dark ultrasound image was unconfirmed.The probe was evaluated with the model 550 phantom and the image was found to be normal.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.The device was serviced, tested, and returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Event Description
Picc rn reported system's ultrasound image is very dark.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIRE-RITE 8 ULTRASOUND SYSTEM 32MM PROBE (REMANUFACTURED)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8067241
MDR Text Key127222172
Report Number3006260740-2018-03213
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138317
UDI-Public(01)00801741138317
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 09/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770601Y
Device Catalogue Number9770601Y
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Event Location Hospital
Date Manufacturer Received09/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-