• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC LINEAR BLUNT PROBE HC; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC LINEAR BLUNT PROBE HC; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9733879
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The active blunt probe was returned to the manufacturer for analysis, and through analysis the issue was able to be duplicated.The returned probe had no color and the leds appeared to be glazed over.When connected to a known good system it was found that led #3 was not firing.Otherwise, the probe returned good geometry and divot error readings.It was determined that there was an electrical failure with the returned probe.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the site had a damaged active blunt probe.One led light of the probe's long cable went out.There was no patient present when this issue was observed.It could not be determined if the instrument passed verification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR BLUNT PROBE HC
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key8067493
Report Number1723170-2018-05647
Device Sequence Number2701825
Product Code HAW
UDI-Device Identifier00613994480613
UDI-Public00613994480613
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9733879
Device Catalogue Number9733879
Device Lot Number150626
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2018
Initial Report FDA Received Date11/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-