According to the reporter during a laparoscopic appendectomy when removing the tissue, the surgeon placed the appendix in the bag, however the appendix was not secured in the bag.The bag was checked and hole was found in the bag.The procedure was completed with another device.The surgical time was extended by less than 30 min.There was no patient harm.
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Distal end of specimen pouch was stretched, deformed and torn before the weld.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur if removal of specimen pouch is attempted through an inadequately sized removal site; there will be pressure exerted on the specimen pouch, usually in the distal end, where there will be stretching and deforming of the specimen pouch until it subsequently rips under tension.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|