Insulin pump passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and dat test at 0.08750 inches.The stop (idle) current and run current measurement tests are within specification.Unit also passed self test, off no power alarm test and a21 error test.No unexpected low battery alarm or unexpected off no power alarm noted during testing.During the occlusion test, the unit recognized the water filled reservoir that was installed in the reservoir compartment.The unit was programmed with a 2.0 unit fill cannula delivery.Then the unit delivered the 2.0 insulin pump properly with water exiting the infusion set.No prime anomaly noted.Insulin pump had intermittent button response on the esc, up arrow and down arrow buttons due to unlocked keypad flex connector on j2 lcd/b.No button error alarm noted.Insulin pump had minor scratched display window, cracked display window, cracked battery tube threads, scratched reservoir tube window and cracked reservoir tube lip.Insulin pump was received without the original battery cap.Unable to verify damage to the battery cap.
|