| Model Number |
MMT-722NAB |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
11/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump alarmed compromised force sensor system and they experienced high blood glucose levels.The customer¿s blood glucose level was 400 mg/dl.Customer treated their high blood glucose levels via manual injection.Customer also reported that the insulin was squirting out during manual prime process.Customer was unable to exit the preparing to prime loop.Customer advised during the seating of the force the sensor did not detect the reservoir.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|