SMITH & NEPHEW, INC. REF HA THREE HOLE 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71336252
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/23/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that revision surgery was performed.Mechanical fall secondary to etoh (alcohol), fell on right side, causing periprosthetic fracture.Afterwards, replacement of shell and liner was performed due to recurrent dislocations.
 
Additional Manufacturer Narrative
The associated complaint devices were not returned.A clinical evaluation was conducted and based on the surgeon¿s statement provided, the root cause of the peri- prosthetic fracture was due to the patient¿s intoxication, trip and fall on her right side.The future impact to the patient beyond the revision cannot be determined.No further clinical/medical assessment is warranted at this time.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of manufacturing records did not reveal any abnormalities that could have contributed to the reported issue.This included a review of sterilization documents which indicated that the product was sterilized according to sterilization release documentation.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF HA THREE HOLE 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8068704
Report Number1020279-2018-02470
Device Sequence Number13842413
Product Code MEH
UDI-Device Identifier03596010452726
UDI-Public03596010452726
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K990666
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/10/2021
Device Catalogue Number71336252
Device Lot Number11HM01699A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/29/2018
Supplement Date Received by Manufacturer10/29/2018
Initial Report FDA Received Date11/14/2018
Supplement Report FDA Received Date02/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-