| Catalog Number |
7205357SR |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the device generate heats during the operation.No patient injury reported.
|
| |
|
Event or Problem Description
|
|
It was reported that before the procedure, device generated heat during operation.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A relationship, if any, between the subject device and the reported event could not be determined.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with motor stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The motor/gearbox assembly could not be removed from the housing for further assessment due to the design of the mdu.The motor/gearbox for this type of mini mdu is potted inside of the housing and unit is scrapped when the motor/gearbox jams or wears out.A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
|
| |
|
Search Alerts/Recalls
|