Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); No Code Available (3191)
Date of Event 10/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pt info: unk.(b)(4).Device evaluation: lens was returned liquid, in a specimen cup.Visual inspection found a torn haptic.(b)(4).
 
Event or Problem Description
The reporter indicated that a 13.2mm, micl13.2, -10.50 diopter, implantable collamer lens was implanted into the patient's left eye (os) on (b)(6) 2018.On (b)(6) 2018 the lens was exchanged for a shorter length lens due to excessive vault and angle closure with elevated iop.This exchange resolved the problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key8068768
Report Number2023826-2018-01674
Device Sequence Number11332542
Product Code MTA
UDI-Device Identifier00841542103305
UDI-Public00841542103305
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Model NumberMICL13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2018
Initial Report FDA Received Date11/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE- SFC-45, LOT#UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
-
-