| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); No Code Available (3191)
|
| Date of Event |
10/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pt info: unk.(b)(4).Device evaluation: lens was returned liquid, in a specimen cup.Visual inspection found a torn haptic.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm, micl13.2, -10.50 diopter, implantable collamer lens was implanted into the patient's left eye (os) on (b)(6) 2018.On (b)(6) 2018 the lens was exchanged for a shorter length lens due to excessive vault and angle closure with elevated iop.This exchange resolved the problem.
|
| |
|