Batch review performed on 14 november 2018: lot 163607: (b)(4) items manufactured and released on 15-sep-2016.Expiration date: 2021-08-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Additional components revised: liner: mpact 01.32.3241hct flat pe hc liner ø32/d, lot 177246: (b)(4) items manufactured and released on 01-mar-2018.Expiration date: 2023-02-04.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Ball heads: mectacer 01.29.205 biolox delta ceramic ball head 12/14 ø 32 size m 0, lot 180534: (b)(4) items manufactured and released on 12-jul-2018.Expiration date: 2023-06-26.No anomalies found related to the problem.To date (b)(4) items of the same lot have been already sold without any similar reported event.
|