• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FUSE; ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FUSE; ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT Back to Search Results
Model Number FSC-3300-ST
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although expected, the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, all images turns black and blue when angulating the scope.The anatomy of the patient was not visible on the center screen.Another fuse 1c colonoscope was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).A fuse 1c colonoscope was received for analysis.A functional evaluation was performed and found that the image flickers when angulated due to a failed charged couple device (ccd).The reported issue was not confirmed.The image flickers when angulated.Therefore, the assigned investigation conclusion code for this complaint is designated as cause traced to component failure.
 
Event or Problem Description
It was reported to boston scientific corporation that a fuse 1c colonoscope was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, all images turns black and blue when angulating the scope.The anatomy of the patient was not visible on the center screen.Another fuse 1c colonoscope was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSE
Common Device Name
ENDOSCOPE, ACCESSORIES, IMAGE POST-PROCESSING FOR COLOR ENHANCEMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8069213
Report Number3005099803-2018-61415
Device Sequence Number1558114
Product Code PEA
UDI-Device Identifier00816849019666
UDI-Public00816849019666
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K160275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFSC-3300-ST
Device Catalogue Number60093
Device Lot NumberF1C0580E1C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Initial Date Received by Manufacturer 10/26/2018
Supplement Date Received by Manufacturer12/04/2018
Initial Report FDA Received Date11/14/2018
Supplement Report FDA Received Date12/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-