| Model Number |
N/A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pt info: requested, not provided.Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.For this reason, evaluation code has been referenced in the conclusions.A review of the device history record of the product code/lot# combination was conducted with no findings.
|
| |
|
Event or Problem Description
|
|
The user facility reported that ik sheath was kinked before coming out of the package.The package looked like it had never been opened.The case was a peripheral leg case.The patient was reported to be in stable condition.The procedure was completed successfully with another sheath.Additional information was received on (b)(6) 2018.A second terumo sheath was opened and used to complete the case successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.2 to provide the completed investigation results.One 5fr sheath and dilator were returned for evaluation.The sheath and dilator were returned in their original manufacturing positions in the original packaging.The packaging had been opened as the seals were not intact.The sheath and the dilator were removed from the original packaging.The sheath was visibly kinked 7.2365" from the distal tip, and, after removal from the sheath, the dilator was visibly kinked 3.3885" from the distal tip.There is no evidence that this event was related to a device defect or malfunction.Based on the investigation results, it is likely that the application of off axis forces may have caused the reported event.However, the exact cause of the reported event cannot be definitively determined based on the available information.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the device return date and update.The actual device has been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.
|
| |
|