| Model Number |
CD3269-40Q |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that an asymptomatic patient presented in clinic during follow-up.Upon review of the stored electrogram, the implantable cardioverter defibrillator had non-sustained right ventricular oversensing episodes due to post-paced t-wave oversensing.Programming changes were made.
|
| |
|