Catalog Number CAT8XTORQ115
Medical Device Problem Codes Device Damaged Prior to Use (2284); Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the returned cat8 had a kink approximately 31.0 cm from the hub and ovalization approximately 60.0 cm from the hub.Conclusions: evaluation of the returned cat8 revealed two kinks along the length of the catheter shaft.These damages were likely a result of forcefully retracting the cat8 from its packaging at extreme angles.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a medical procedure, the hospital staff noticed that an indigo system aspiration catheter 8 (cat8) had two major kinks before removal from its packaging hoop.The damage to the cat8 was found prior to use and therefore, it was not used in the procedure.The procedure was completed using a new cat8.There was no report of adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 8
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8069383
Report Number3005168196-2018-02255
Device Sequence Number16533352
Product Code DXE
UDI-Device Identifier00814548016290
UDI-Public00814548016290
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/17/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/27/2021
Device Catalogue NumberCAT8XTORQ115
Device Lot NumberF82567
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2018
Initial Report FDA Received Date11/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-