• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Medical Device Problem Code Component or Accessory Incompatibility (2897)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that during biomed testing, the device was unable to detect the attached multifunction cable.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
 
Additional Manufacturer Narrative
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also correcting information submitted on the initial medwatch report.The device was evaluated by zoll medical corporation.The reported malfunction was duplicated and attributed to a faulty spo2 connector.The spo2 connector was replaced to remedy the reported problem.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.This report was inadvertently submitted and reports of this nature are typically not submitted as there would be no potential for clinical impact.
 
Event or Problem Description
Complainant alleged that during biomed testing, the device was unable to detect the attached spo2 cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key8069448
Report Number1220908-2018-03313
Device Sequence Number14023301
Product Code MKJ
UDI-Device Identifier00847946002572
UDI-Public00847946002572
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K060559/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2018
Supplement Date Received by Manufacturer10/25/2018
10/25/2018
Initial Report FDA Received Date11/14/2018
Supplement Report FDA Received Date12/07/2018
02/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-