| Model Number |
R SERIES |
| Medical Device Problem Code |
Component or Accessory Incompatibility (2897)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device was unable to detect the attached multifunction cable.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also correcting information submitted on the initial medwatch report.The device was evaluated by zoll medical corporation.The reported malfunction was duplicated and attributed to a faulty spo2 connector.The spo2 connector was replaced to remedy the reported problem.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.This report was inadvertently submitted and reports of this nature are typically not submitted as there would be no potential for clinical impact.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device was unable to detect the attached spo2 cable.
|
| |
|