WRIGHTS LANE SYNTHES USA PRODUCTS LLC 13.0MM CANNULATED DRILL BIT 300MM; INSTR,SURGICAL,ORTHOPEDIC,AC-POWERED MOTOR/ACCESS & ATTACH
|
Back to Search Results |
|
| Model Number |
351.27 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional product codes gfa, gff, hsz.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2018, a cannulated drill bit became stuck in the guide sleeve and would not rotate during an unknown procedure.Both items were removed from the surgery.The nurse took them back to sterile processing to be cleaned by the technician.After cleaning, the drill bit was still stuck and not rotating.There was no surgical delay reported.Patient and procedure outcome are unknown.This report is for one (1) 13.0mm cannulated drill bit 300mm.This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device was received, and the product evaluation is in progress.No conclusion can be drawn.Additionally, device history records review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Dhr review was completed.Part number: 351.27.Synthes lot number: u224833.Supplier lot number: u224833.Release to warehouse date: 08-jun-2015.Manufacturing site: synthes monument.Supplier: orchid unique.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Customer quality conducted an investigation of the returned device.Investigation flow: functional: visual inspection: visual inspection confirmed the reported complaint condition as the device was returned stuck in its mating device (protection sleeve).The device could not be disassembled.No further issues were identified.The complaint condition replication and functional inspection is not possible due to the device returned in stuck condition.The received condition does agree with the complaint description.The complaint is confirmed.Dimensional inspection: cannulated 3- fluted drill with large quick coupling.The above drawing was reviewed during the investigation and no design issues were identified.With major portion of the device being stuck in protection sleeve and with complexity in geometry of the helical flutes at the distal portion accurate measurement on the device both at relevant and accessible drill flute portions could not be performed.Document/specification review: review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Conclusion: no definitive root cause could be determined from the provided information, however during visual inspection a very slight off-axis insertion of drill was observed.Therefore, it is possible that any off-axis alignment could have likely contributed to the complaint condition.The complaint is confirmed.However, no design or manufacturing defect or deficiency was observed during the investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|