MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE ICONIX 1 NEEDLES WITH INTELLIBRAID TECHNOLOGY, 1.4MM ANCHOR WITH 1.2MM XBRAID TT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE ICONIX 1 NEEDLES WITH INTELLIBRAID TECHNOLOGY, 1.4MM ANCHOR WITH 1.2MM XBRAID TT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 3910500212
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 10/17/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for investigation therefore the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: drill bit broke.Probable root cause: design: drill design too weak; drill material too weak; manufacturing processes selected weaken drill material; drill extends too far beyond guide ; bushing does not keep drill in orientation with guide (iconix without needles disposable drill only); process: drill manufactured out of specification; application: excessive force while drilling; selection of wrong drill; reuse of disposable drill; use of reusable drill past lifetime (after it has become dull); starting and stopping drill; pulling drill out of bone without running drill; moving drill guide during drilling.
 
Event or Problem Description
It was reported that the drill bit broke.The broken piece remained inside the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICONIX 1 NEEDLES WITH INTELLIBRAID TECHNOLOGY, 1.4MM ANCHOR WITH 1.2MM XBRAID TT
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8069629
Report Number0002936485-2018-00688
Device Sequence Number1928988
Product Code MBI
UDI-Device Identifier37613327283441
UDI-Public37613327283441
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K151201
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910500212
Device Lot Number17297AE2
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2018
Initial Report FDA Received Date11/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age29 YR
-
-