MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number TEXP120RUH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/17/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) patient underwent a breast reconstruction with an artoura breast tissue expander 535cc.Deflation was noticed intraoperatively.The tissue expander would not hold suction and the procedure was delayed 45 minutes.As a result, the device was removed and replaced.There was no patient consequence.
 
Additional Manufacturer Narrative
Device evaluation summary: upon receipt by mentor, the device returned with blue fluid.No foreign material was observed within the device or on the shell surface.Upon visual examination the device appears intact.Leak testing was performed in accordance with mentor procedures and no leak sites were detected.No other anomalies were discovered.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, mentor product analysis lab was precluded from further evaluating the device.The complaint was not confirmed.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key8069966
Report Number1645337-2018-06769
Device Sequence Number19230786
Product Code LCJ
UDI-Device Identifier00081317009542
UDI-Public00081317009542
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date06/24/2022
Device Catalogue NumberTEXP120RUH
Device Lot Number7593993
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Initial Date Received by Manufacturer 10/18/2018
Supplement Date Received by Manufacturer11/20/2018
Initial Report FDA Received Date11/14/2018
Supplement Report FDA Received Date12/11/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-