| Catalog Number |
TEXP120RUH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/17/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) patient underwent a breast reconstruction with an artoura breast tissue expander 535cc.Deflation was noticed intraoperatively.The tissue expander would not hold suction and the procedure was delayed 45 minutes.As a result, the device was removed and replaced.There was no patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: upon receipt by mentor, the device returned with blue fluid.No foreign material was observed within the device or on the shell surface.Upon visual examination the device appears intact.Leak testing was performed in accordance with mentor procedures and no leak sites were detected.No other anomalies were discovered.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, mentor product analysis lab was precluded from further evaluating the device.The complaint was not confirmed.There is no evidence that the issue is related with manufacturing.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|