ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751763
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Eye Burn (2523)
Event Date 10/29/2018
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that a patient suffered a corneal wound burn during phacoemulsification.The doctor felt that either the phaco tip was not on correctly or something was wrong with the irrigation.The patient is healing well.The doctor has requested a company representative visit the facility for further training with staff and troubleshoot on how this event might of happen.
 
Manufacturer Narrative
The finished good lot specific to this event is not known; therefore, lot history and device history record review was not possible.The customer did not retain a cassette sample for this complaint report; visual inspection or functional testing could not be conducted in order to ascertain the failure mode of the consumable device.We could not determine with the available information the source of the potential malfunction; whether it originated from the tip, the cassette, or the handpiece (probe).The root cause of the customer's complaint could not be established as a sample has not been received.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.After review of this complaint, it has been determined that no action will be taken at this time as a sample was not returned and no root cause could be established.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Consumables manufacturing management has been made aware of this complaint through the consumer affairs review meeting.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
cindy milam
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152231
MDR Report Key8070303
MDR Text Key127123266
Report Number2028159-2018-02405
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/14/2018
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751763
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CENTURION OZIL HANDPIECE; CENTURION SURGICAL PROCEDURE PAK; PHACOEMULSIFICATION SURGICAL TIP
Patient Outcome(s) Other;
Patient Age83 YR
-
-