| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); Confusion/ Disorientation (2553)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
The reporter contacted animas on (b)(6) 2018 alleging the patient experienced hypoglycemia while on insulin pump therapy evidenced by blood glucose measuring 48 mg/dl with confusion and lightheadedness.The patient reportedly did not receive any treatment above or beyond the usual routine of diabetes care and management.The reporter alleged a history/settings issue with the pump; the basal history reportedly does not match the active basal program.This complaint is being reported because the patient reportedly experienced serious injury while on insulin pump therapy associated with a history/settings issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 25-mar-2019 with the following findings: a review of the black box showed manual time changes occurred.The daily insulin delivery totals appeared inconsistent due to the time/date changes.The pump passed delivery accuracy testing and was found to be delivering within the required specifications.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|