• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: X-SPINE SYSTEMS, INC. AXLE INTERSPINOUS FUSION SYSTEM; INTERSPINOUS PROCESS FIXATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

X-SPINE SYSTEMS, INC. AXLE INTERSPINOUS FUSION SYSTEM; INTERSPINOUS PROCESS FIXATION SYSTEM Back to Search Results
Model Number X060-0320
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complainant reported that a screwdriver broke while removing a crosslink from a previous surgery.There were no known patient complications.The distal tip of the driver was sheared off and the remaining portion of the star tip was twisted.Excess rotational force may cause the start tip to rotate.The instrument was reported to be used to remove a crosslink from a previous surgery.The complaint instrument is not intended to remove crosslinks, and this may have contributed to the malfunction.The dhr for the complaint instrument was reviewed and no manufacturing anomalies were identified.The device met all of the required specifications prior to being released to distribution.
 
Event or Problem Description
The complainant reported that a screwdriver broke while removing a crosslink from a previous surgery.There were no known patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXLE INTERSPINOUS FUSION SYSTEM
Common Device Name
INTERSPINOUS PROCESS FIXATION SYSTEM
Manufacturer (Section D)
X-SPINE SYSTEMS, INC.
664 cruiser lane
belgrade MT 59714
MDR Report Key8070711
Report Number3005031160-2018-00032
Device Sequence Number2190086
Product Code HXX
UDI-Device IdentifierM697X06003201
UDI-PublicM697X06003201
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberX060-0320
Device Lot Number057362
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2018
Initial Report FDA Received Date11/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-