UMANO MEDICAL INC. OOK SNOW; MEDICAL BED Back to Search Results
Model Number FL36
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/17/2018
Type of Reportable Event Malfunction
Event or Problem Description
User reported to sales representative that a bed exit system on a medical bed detects patient exiting intermittently.
 
Additional Manufacturer Narrative
No defects were found.The most probable cause of this situation appears to be that the calibration had not been done by the user and the instructions for zeroing not followed.Therefore, this would be a user error since the indications for calibration and zeroing are mentioned in the accompanying documentation.Precisions were made to the user manuel (corrective action : res 85592 recall 3009591865-04/28/2020-001-c ) to emphasis the preventive maintenance annual checklist, which includes calibration and to emphasis on usage of detection system and the zeroing procedure, which is explained and described in the user manual.The manufacturer will follow the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OOK SNOW
Common Device Name
MEDICAL BED
Manufacturer (Section D)
UMANO MEDICAL INC.
230 boulevard nilus-leclerc
l'islet, G0R 2 C0
CA  G0R 2C0
MDR Report Key8070746
Report Number3009591865-2018-00017
Device Sequence Number13991859
Product Code FNL
UDI-Device Identifier00670482000104
UDI-Public010067048200010411180711
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Remedial Action Inspection,Modification/Adjustment,Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFL36
Device Catalogue NumberFL36
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/17/2018
Supplement Date Received by Manufacturer10/17/2018
10/17/2018
Initial Report FDA Received Date11/14/2018
Supplement Report FDA Received Date12/14/2018
12/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight36
-
-