Catalog Number 151640906
Medical Device Problem Codes Difficult to Insert (1316); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Atunne femoral component 9 and tibia component 7 were used, when they tried to implant the attune insert size 9/6mm the insert did not sit on the tibial plato, and appeared to have an overhang and the knee could not function properly.They then used the attune insert 9/5mm which fit perfectly with no problem.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary:no device was returned.Depuy synthes considers the investigation closed at this time.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: the device was reviewed by the manufacturing site and concluded that the issue was not manufacturing related.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary
=
> the device was reviewed by the manufacturing site and concluded that the issue was not manufacturing related root cause undetermined depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.The device was reviewed by the manufacturing site raynham and a report was received stating; manufacturing review of the dhr associated with the manufacturing of p/n 151640906 (batch c54654) along with physical examination of the part identified that the issue was not manufacturing related.The returned part was evaluated by quality engineering and process engineering and damage was observed on the inferior face of the part.The inferior snap tab is protruding in the z direction due to its placement, and subsequent removal, from the tibial plateau as the surgeons swapped this attune insert 9/6mm with an attune insert 9/5mm.The protrusion of the tab makes dimensional confirmation testing impossible because the flat face of the inferior side of the insert cannot be used as a reliable basis of coordination by the cmm.Dimensional confirmation testing would not be representative of the actual dimensions of this part, so cmm verification would lead to undue alarm.Dhr review: the dhr showed that all 12 parts in batch c54654 passed cmm inspection and provided a sample of cmm dimension measurements used in the acceptance process.In process control: there is a 100% verification of all ctqs for 100% of products in a batch.A worldwide complaint search was conducted for complaints related to product # 151640906.The list was reviewed for instances of dimensionally nonconforming attune inserts.The review returned no instance of this no information received with this individual complaint indicated that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.The complaint shall be closed with an undetermined conclusion; entered into the complaints database and monitored through trend analysis.Post market surveillance is per (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FB INSRT SZ 9 6MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
MDR Report Key8070756
Report Number1818910-2018-75497
Device Sequence Number2154449
Product Code JWH
UDI-Device Identifier10603295050490
UDI-Public10603295050490
Combination Product (Y/N)N
Initial Reporter CountrySI
PMA/510(K) Number
K111433
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number151640906
Device Lot NumberC54654
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2019
Initial Date Received by Manufacturer 10/24/2018
Supplement Date Received by Manufacturer01/22/2019
07/01/2019
07/01/2019
Initial Report FDA Received Date11/14/2018
Supplement Report FDA Received Date01/22/2019
07/08/2019
07/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-