• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH SI POLYAXL SCREW 7 X 35MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH SI POLYAXL SCREW 7 X 35MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 179712735
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The l5 lumbar spondylolisthesis reconstruction surgery was performed with expedium si and hook system on (b)(6) 2018.It was reported that the surgeon could not apply the torque for final tightening the set screws.The surgeon confirmed there was no cross-threaded, so that the surgeon attempted to tighten couple of time, but the surgeon could not tighten the set screws.Then, the surgery was completed by replacing the screw which removed from the patient¿s lumbar.After the removing the screws, the surgeon recognized there was metal like fragment around the screw.There was 45min surgical delay and there was no adverse consequence to the patient.No further information was provided by the hospital.
 
Additional Manufacturer Narrative
(b)(4).Visual examination found that the tulip head¿s threads were torn off.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.With the information provided, a definitive root cause for the torn threads on the si polyaxial screw 7 x 35mm cannot be determined.Noted damage suggests that inadvertently cross threading of the setscrew occurred upon insertion into the tulip head.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).Date of awareness for follow up # (b)(4) should have been 11/20/2018 not 11/2/2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SI POLYAXL SCREW 7 X 35MM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key8071201
Report Number1526439-2018-51067
Device Sequence Number19221913
Product Code NKB
UDI-Device Identifier10705034067889
UDI-Public(01)10705034067889
Combination Product (Y/N)N
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number179712735
Device Catalogue Number179712735
Device Lot NumberAHDBK1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Initial Date Received by Manufacturer 11/02/2018
Supplement Date Received by Manufacturer11/02/2018
01/03/2019
Initial Report FDA Received Date11/14/2018
Supplement Report FDA Received Date12/03/2018
01/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-