• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. VALEO BALLOON EXPANDABLE STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. VALEO BALLOON EXPANDABLE STENT Back to Search Results
Model Number EV07261CD
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 04/30/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 05/2019), (manufacturing date: 05/2017).
 
Event or Problem Description
It was reported through the results of a clinical trial that approximately six months post deployment of a balloon expandable stent in the right superficial femoral artery, duplex ultrasound demonstrated in-stent restenosis in the target lesion.Approximately eight months post stent deployment, duplex ultrasound demonstrated the target lesion was occluded.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only event reported to date for this lot number and failure mode.Investigation summary: the investigation is inconclusive for the reported occlusion, as the device was not returned for evaluation.The definitive root cause for the reported occlusion could not be determined based upon available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported through the results of a clinical trial that approximately six months post deployment of a balloon expandable stent in the right superficial femoral artery, duplex ultrasound demonstrated in-stent restenosis in the target lesion.Approximately eight months post stent deployment, duplex ultrasound demonstrated the target lesion was occluded.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALEO BALLOON EXPANDABLE STENT
Common Device Name
BALLOON EXPANDABLE STENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8072370
Report Number2020394-2018-02032
Device Sequence Number1558147
Product Code FGE
UDI-Device Identifier00801741000737
UDI-Public(01)00801741000737
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K052132
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEV07261CD
Device Catalogue NumberEV07261CD
Device Lot NumberGFBR0702
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/18/2018
Supplement Date Received by Manufacturer01/15/2019
Initial Report FDA Received Date11/14/2018
Supplement Report FDA Received Date02/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age73 YR
-
-