LIFESCAN EUROPE GMBH OT VERIO TEST STRIPS; GLUCOSE MONITORING SYS/KIT Back to Search Results
Lot Number 4353954
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Event or Problem Description
Device evaluation: the lay user/patients test strips have been returned and evaluated by lifescan product analysis with the following findings: the test strips failed testing; the reported issue was confirmed.The test strips were found to be sticking together due to static or small amounts of adhesive.Complaints relating to the reported product was evaluated.It was concluded that the number of complaints for the product did not breach thresholds indicative of a systemic issue.On (b)(6) 2018, the lay user/patient contacted lifescan usa, alleging that their ot verio test strips were sticking together.This complaint was initially ruled out because during customer services troubleshooting, there was no indication to reasonably suggest that the product malfunctioned and there was also no allegation of an adverse event as a result of the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO TEST STRIPS
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE GMBH
gubelstrasse 34
zug 6300
SZ  6300
MDR Report Key8072825
Report Number3008382007-2018-03133
Device Sequence Number11352270
Product Code LFR
UDI-Device Identifier00353885009751
UDI-Public00353885009751
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Lot Number4353954
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date11/12/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2018
Initial Report FDA Received Date11/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-