| Catalog Number |
NCEUP2515X |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An attempt was made to use one nc-euphora rx ptca balloon catheter to treat a moderately tortuous and calcified lesion exhibiting 90% stenosis located in the mid obtuse marginal (om).The device was inspected with no issues noted.Negative prep was performed with no issues.The lesion was not pre-dilated.The device passed through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.It was reported that a balloon leak occurred on the first inflation while inflating to 12atm.It is indicated that there was a hole in the balloon prior to inflating the device.The patient is reported to be alive with no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The balloon was intended to be used for pre-dilation.The device was not moved or re-positioned prior to the inflation difficulties.The same inflation device was successfully used with other devices before and after the inflation difficulties were encountered.Another balloon of the same size was used to complete the procedure.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the device returned with crystallised contrast visible in the balloon and inflation lumen.The device was placed in the waterbath to disperse the contrast in order to aid inflation.The balloon failed negative prep.On pressurisation of the device, liquid was observed exiting the balloon distal cone.The balloon failed to maintain pressure.Upon visual inspection of the device, there was a short longitudinal tear on the balloon material at the distal cone.There was no lead in scratches visible surrounding the tear site.The material at the tear site was protruding outwards.No other damage evident to the remainder of the device.If information is provided in the future, a supplemental report will be issued.
|
| |
|